RPANEL is a software that allows remote control over R series scales, MAS and MPS modules. 

RPANEL can automatically detect to which COM port the scale or module is connected. It can also, to some extent, detect baudrate; however, this may fail in some circumstances. In is then required to modify the device's baudrate manually using the terminal. This option is not available for MAS and MPS modules if the device operates without a terminal.

In such situation, the operator is stuck with no possibility to connect the device to RPANEL.


BUT!


----------------------------------------------------------------------------------------------------------------------------------


How to change baudrate of R series scale, MAS and MPS modules when having no access to buttons/terminal. 


1. download and run a software terminal tool (such as Hercules) on your computer,

2. connect to the device via its currect connection parameters (COM port and baudrate)

3. use command 


DGJL699541$0D$0A 


to enable factory level access on the device

4. use command


CSPE X,Y$0D$0A 


to change the baudrate of the device where X stands for:

1 - change baudrate of COM 1

2 - change baudrate of COM 2

and where Y stands for various baudrates:

1 - 4800

6 - 115200


Example 1: CSPE 1,6$0D$0A - will change COM 1 to baudrate 115200

Example 2: CSPE 2,1$0D$0A - will change COM 2 to baudrate 4800


5. after using both commands disconnect from the device and restart it